Instruction Manual
EL 9000 A Series
EN
Date: 11-16-2011
A changeover to regulation mode 
CR can reset the set 
values of resistance, current and power to their nominal 
values, if the parameter 
Keep set values
 has been set 
to 
no
 in the setup. If set to 
yes
, the least adjusted set 
6.7  Usage of Level A and Level B
Introduction
The terms Level A and Level B stand for two different 
sets of set values which can be switched over in order 
to generate a value step. This is either done manually 
with the selector 
Level Control (3) or externally via the 
analogue  interface  with  the  trigger  input  (only  in 
A/B
mode) or automatically (in 
A/B mode).
Each of them, 
A and B, has five set values dedicated to 
the four regulation modes. That means, that the set value 
of current is dedicated to constant current mode etc. In 
CP mode you can, for example, adjust two set values for 
power, switch between them and generate power steps. 
When using the A/B mode (see 6.7.3), this switchover 
happens automatically, in combination with the adjustable 
pulse widths (i.e. pulse times) of A and B. This generates 
a  square  wave  shaped  set  value,  whose  high  level  is 
represented by the A value and whose low level by the 
B value and whose period time (and thus frequency) by 
the sum of the variable pulse widths of A and B. These 
also  determine  the  so-called  duty  cycle. An  example: 
A = 10ms, B = 90ms, this results in a period time of 100ms 
(= 10Hz) with 10% duty cycle.
6.7.1  Level A
When  switching  to  „Level A“  with  the  selector 
Level
Control(3),  the  set  values  of  Level A  become  active. 
This is only possible, if the status is not 
Remote mode
 
or 
External mode
. After switching to 
Level A mode the 
display shows 
Level A.
 An arrow (->) is placed next to 
the set value which is dedicated to the currently chosen 
regulation  mode,  in  order  to  be  instantly  adjustable. 
The set value(s) can be selected with 
Selection(5) and 
changed with 
Setting(6). Which set values are selectable 
depends  on  the  regulation  mode.  In  constant  current 
mode these are, for example, the current and the power, 
because the set value for resistance is only available in 
resistance mode and the voltage value must be set to zero 
in constant current mode. See figure 9 on the next page.
The set values of this mode are kept unchanged until 
they are adjusted again, even if the load is switched to 
Level BA/B or Battery test. This does not apply if the 
parameter 
Keep  set  values
  has  been  set  to 
no
  (see 
„7.1. The setup menu“). Then the set values are reset to 
default values, depending on the preselected mode (also 
Level A mode, the load can be switched to remote 
control from a PC and via the interface, in order to control 
and monitor it like during manual operation.
When  changing  to  remote  control  the  currently  selec-
ted 
Level Control setting is retained and can then be 
changed  by  a  command,  except  with  GPIB  interface 
IF-G1.
Handling the device 
A changeover to regulation mode 
CV can reset the set 
values  of  voltage,  power  and  current  to  their  nominal 
values, if the parameter 
Keep set values
 has been set 
to 
no
 in the setup. If set to 
yes
, the least adjusted set 
values are kept. Also see „7.1. The setup menu“.
Note: the CV regulation can not be used in combination 
with the battery test mode. Selecting it for battery test will 
indicate an error message on the display.
Note:  the  voltage  set  value,  which  is  assigned  to  the 
regulation mode CV, must be set to 0 in any other regu-
lation mode. That‘s why it is not accessible in the manual 
operation in the CC, CP or CR modes. 
During 
remote control the voltage set value can be sent 
to the device, but will be ignored and an access error will 
be reported by the communication.
During 
external control (analogue interface) there is an 
exception: the voltage set value must be given and can 
be used as desired, but should be set to 0V if normal CC, 
CP or CR operation is required.
CP
 means constant power. Here are the set values of 
power and current adjustable. In this mode the device 
withdraws  as  much  current  as  neccessary  from  the 
feeding source in order to consume the adjusted power, 
dependant on the input voltage (P = U * I). In case the 
input voltage is so low that the withdrawn current exceeds 
the nominal current of the device, the adjusted power can 
not be achieved. Simultaneously, the current limitation 
becomes active. 
A changeover to regulation mode 
CP can reset the set 
value of current to its nominal value and the set value of 
power to 0, if the parameter 
Keep set values
 has been 
set to 
no
 in the setup. If set to 
yes
, the least adjusted set 
CR
  means  constant  resistance. The  device  features 
two resistance ranges. For values refer to the technical 
specifications section.
The CR regulation mode is only active as long the switch 
Mode(2) is set to one of the two resistance ranges. Both 
ranges act the same way, they only differ in the higher 
resolution and accuracy of the lower resistance range. 
In this mode, the set values of resistance, current and 
power are adjustable.
This regulation mode loads the feeding voltage or current 
source so much that the adjusted resistance results from 
the relation of input voltage to input current (R= U / I). In 
case the input voltage is so high that the input current ex-
ceeds the nominal current value, the adjusted resistance 
can not be achieved. 
Simultaneously, the current limitation becomes active. If 
the product of input voltage and input current,  i.e. the po-
wer, exceeds the nominal power of the device, the power 
limitation becomes active. Then the adjusted resistance 
can also not be achieved.