6.  Handling
For an overview of all operating elements also see sec-
6.1  The display
This is an overview of the two line display and its layout. 
The left side always shows actual values while the load 
input is  switched on:
Figure 3
The 
indicator for the regulation mode(a triangle) appears 
next  to  the  actual  value,  which  is  related  to  currently 
active regulation mode of the load. This can differ from 
the chosen regulation mode if a nominal value has been 
exceeded. The current or power limitation have priority 
over the voltage or resistance limitation here. That means, 
that the load will change into constant current limitation as 
soon as the nominal current value has been exceeded. 
The triangle indicates this.
In standby (load input off), the status 
Standby
is displayed 
and only the actual value of voltage is shown:
Figure 4
Various texts in the right half of the display indicate diffe-
rent operation modes or errors:
Figure 5
This text appears if the load was set to remote control 
operation (
Remote mode
) via the optional interface card. 
This is possible in the modes 
Level AB and A/B.
Figure 6
The  text 
External  mode 
indicates,  that  the  control  of 
the  load  was  switched  to  the  analogue  interface. The 
set values can not be adjusted at the device anymore in 
this mode. For detailed information see section „8. The 
Handling the device
Alarm management
The  signals 
Overvoltage
or 
Power  fail
indicate  the 
presence of an error.
power fail alarm occurs if the mains input voltage is 
too low. An
overvoltage alarm is indicated if the overvol-
tage limit for the DC input has been exceeded (see „2.2. 
Device specific data“ for the limit). Both alarms switch 
the load input off. After removal of the cause it can be 
put online again.
Alarm  indications  remain  in  the  display  until  acknow-
ledged. The purpose is to inform the user that an alarm 
is either active or already gone. Example:
Figure 7
An active alarm has priority over a gone alarm. If two alarms 
are active, then 
overvoltage has priority over power fail 
in the display.
Only alarms with status „(gone)“ can be acknowledged. 
Acknowledging is either done with the pushbutton „Input 
on/off“ or by reading the internal alarm buffer via a digital 
interface card. The buffer is cleared after reading, if no 
alarm is currently active.
In analogue remote control, alarms also have to be ack-
nowledged by pulling pin REM-SB to low, i.e. switching 
the input off. If the input is already off while the alarm 
occurs, it will automatically acknowledge an alarm with 
status „(gone)“. 
Note: with interface card GPIB plugged, the internal alarm 
buffer is continuously read by the card and will thus au-
tomatically acknowledge gone errors, so these are not 
displayed anymore. Via SCPI command, errors with status 
„(gone)“ are not returned, only active errors. When using 
an Ethernet card, alarms like these are not returned at all.