Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 Scheda Tecnica

Codici prodotto
GSP-830
Pagina di 181
  
TRIGGER 
   105
 
Activate external trigger 
 
Panel operation 
1.  Press the Trigger key. 
Trigger
 
 
2.  Press F2 (Trigger Condition 
Ext.). 
Trigger
Condition
Video  Ext.
F   2
 
 
3.  The Ext. trigger icon appears at 
the bottom of  the display. 
 
 
4.  Connect the external trigger 
signal to the rear panel terminal.
EXT. TRIG.
INPUT
 
Input level range 
0V to 5V, positive edge trigger 
Select Trigger Mode 
 
Panel operation 
1.  Press the Trigger key. 
Trigger
 
 
2.  Press F3 (Trigger Mode) 
repeatedly to select the trigger 
mode. 
Trigger Mode
Norm. Sgl. Cont.
F   3
 
3.  Press F6 (Run Now) to 
manually start triggering. 
Run Now
F   6
 
Modes Normal 
The GSP-830 captures signals every time 
the triggering condition occurs. 
 Single 
The GSP-830 captures signals when the 
first triggering condition occurs, then 
stops capturing altogether. 
 Continuous 
The GSP-830 captures signals when the 
first triggering condition occurs, then 
switches to Free Run mode. After then, 
input signals are captured continuously 
regardless of  their conditions.