Making Rise Time Measurements 
To measure rise time, do the following: 
CURSOR
 
From scope mode, display the 
cursor key labels. 
 
F1
 
Press to highlight   (rise time). 
Observe that two horizontal 
cursors are displayed. 
F4
 
For multiple traces select the 
required trace A, B, C, D or M (if a 
math function is active). 
F3
 
 Select MANUAL or AUTO (this 
automatically does steps 5 to 7).  
 
Move the upper cursor to 100% of 
the trace height. A marker is 
shown at 90%. 
F2
 
Highlight the other cursor. 
 
Move the lower cursor to 0% of 
the trace height. A marker is 
shown at 10%. 
The reading shows the risetime from 10%-90% of the 
trace amplitude. 
 
Figure 27. Risetime Measurement