Lecroy WAVEACE 2034 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2034 Manuale Utente

Codici prodotto
WaveAce 2034
Pagina di 89
Operator's Manual 
920836 Rev A  
55 
Bit 
Value 
Name 
Description 
Note 
DDE 
1 - Device Dependent 
(specific) Error occurred. 
The Device specific Error (DDE) is set true (1) whenever 
a hardware failure has occurred at power-up or at 
execution time, such as a channel overload condition, 
or a trigger or timebase circuit defect. The origin of the 
failure can be localized with the DDR? query. 
QYE 
1 - QuerY Error occurred.  The QuerY Error bit (QYE) is set true (1) whenever (a) 
an attempt is made to read data from the Output 
Queue when no output is either present or pending, (b) 
data in the Output Queue has been lost, (c) both 
output and input buffers are full (deadlock state), (d) 
an attempt is made by the controller to read before 
having sent an <END>, (e) a command is received 
before the response to the previous query was read 
(output buffer flushed). 
RQC 
0 - Oscilloscope never 
ReQuests bus Control. 
The ReQuest Control bit (RQC) is always false (0), as the 
oscilloscope has no GPIB controlling capability. 
OPC 
0 - OPeration Complete 
bit not used. 
The OPeration Complete bit (OPC) is set true (1) 
whenever *OPC has been received, since commands 
and queries are strictly executed in sequential order. 
The oscilloscope starts processing a command only 
when the previous command has been entirely 
executed 
STATUS - *OPC 
Description 
The *OPC (OPeration Complete) command sets to true the OPC bit (bit 0) in the Standard Event 
Status Register (ESR). 
The *OPC? query always responds with the ASCII character 1 because the oscilloscope only 
responds to the query when the previous commands are entirely executed. Since *OPC? 
executes only when processing is complete, it is recommended for synchronization purposes.   
Command Syntax 
*OPC 
Query Syntax 
*OPC? 
Response Format 
*OPC 1