1. Important
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and 
SYTEMS:
7KHPRQLWRULVLQWHQGHGIRUXVHLQWKHHOHFWURPDJQHWLFHQYLURQPHQWVSHFLÀHGEHORZ7KHFXVWRPHURUWKHXVHURI
the monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – 
guidance
Electrostatic 
discharge (ESD) 
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or 
FHUDPLFWLOH,IÁRRUVDUHFRYHUHGZLWK
synthetic material, the relative humidity 
should be at least 30%.
Electrical fast 
transient/burst  
IEC 61000-4-4
2 kV for power supply 
lines
1 kV for input/output 
lines
2 kV for power supply 
lines
1 kV for input/output 
lines
Mains power quality should be that 
of a typical commercial or hospital 
environment.
Surge 
IEC 61000-4-5
1 kV line(s) to line(s) 
2 kV line(s) to earth
1 kV line(s) to line(s)
2 kV line(s) to earth
Mains power quality should be that 
of a typical commercial or hospital 
environment.
interruptions and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11 
<5 % UT (>95 % dip in 
UT) for 0.5 cycle
40 % UT (60 % dip in 
UT) for 5 cycles
70 % UT (30 % dip in 
UT) for 25 cycles
<5 % UT (>95 % dip in 
UT) for 5 sec
<5 % UT (>95 % dip in 
UT) for 0.5 cycle
40 % UT (60 % dip in 
UT) for 5 cycles
70 % UT (30 % dip in 
UT) for 25 cycles
<5 % UT (>95 % dip in 
UT) for 5 sec
Mains power quality should be that 
of a typical commercial or hospital 
environment. If the user of the monitor 
requires continued operation during 
power mains interruptions, it is 
recommended that the monitor be 
powered from an uninterruptible power 
supply or a battery.
Power frequency 
(50/60 Hz) 
PDJQHWLFÀHOG
IEC 61000-4-8
3 A/m
3 A/m
3RZHUIUHTXHQF\PDJQHWLFÀHOGVVKRXOG
be at levels characteristic of a typical 
location in a typical commercial or 
hospital environment.
  Note
UT is the a.c. mains voltage prior to application of the test level.