5-2. Attenuator Probes 
 
When measuring signals at input A that will exceed its maximum input voltage 
level (See FIG. 1), a 10:1 attenuating probe such as an oscilloscope probe should 
be used.   
 
NOTE 
DO NOT USE A 10:1 PROBE WITH THE INPUT C. THESE PROBES ARE 
DESIGNED WITH A 10:1 ATTENUATION FOR AN INSTRUMENT WITH AN 
INPUT RESISTANCE OF 1 M 
Ω. THE 50 Ω TERMINATION OF INPUT C WOULD 
RESULT IN UNACCEPTABLY HIGH ATTENUATION. 
 
5-3. Line Frequency Measurements 
  When making line frequency measurements, use of the LOW PASS FILTER, the 
ATTENUATOR, and/or the 10:1 probe is recommended.   
        
WARNING 
USE CAUTION WHEN MEASURING THE LINE FREQUENCY OF AN AC 
OUTLET. USE THE PROBE TIP ONLY, AND MEASURE BOTH SIDES OF THE 
LINE. THE NEUTRAL SIDE WILL GIVE A ZERO READING AND THE HOT SIDE 
WILL PROVIDE THE DESIRED MEASUREMENT. DO NOT USE THE "GROUND" 
LEAD OF THE PROBE. REMEMBER THAT THE CHASSIS OF THE COUNTER 
AND THE “GROUND” LEAD OF THE PROBE ARE ALREADY AT EARTH 
GROUND (VIA THE 3-WIRE POWER CORD OF THE INSTRUMENT). TOUCHING 
THE “GROUND" LEAD TO THE 'HOT" SIDE OF THE LINE WOULD PLACE A 
DIRECT SHORT ON THE POWER LINE THROUGH THE PROBE CABLE, 
RESULTING IN A BIG SPARK, DAMAGE TO THE PROBE CABLE, AND 
POSSIBLE INJURY TO THE USER.