When you measure the output voltage or short current by the multi-meter, please make 
sure the value is correct. Otherwise it will cause the calibration failure and let the 
output turbulence. 
 
Please  make sure that the maximum current of your multi-meter is higher than the 
maximum current of the calibrated power supply, otherwise it would damage the 
multi-meter. 
 
 
 
5.4 Reset to the default calibration settings 
In the calibration process, if there is any failure caused by the misreading the voltage value and current 
value or any other reason, you can use the “Restore” to let the power supply return to the state before 
calibrating. 
 
 
 
 
 
 
 
 
 
 
 
Chapter 6 PV1785B_1788 Software 
6.1 Introduction 
Software PV1785B_1788 is control software for 1785B - 1788 series programmable power supplies. It 
can work with all  single-output power supply models. Please make sure that you purchased IT-E131 
communication cable, and use IT-E131 to connect the power supply and computer before the 
communicating.    This  software can  accomplish  all the functions of the power supply, such as setting 
constant voltage, setting constant current, max voltage etc. Also you can make quickly settings or make 
a program for the output voltage and current by the computer.  In addition,  PV1785B_1788 offer  a 
GO/NG test function for automatic factory testing  .It brings you much convenient when you use the 
power supply. 
 
 
6.2 Installation 
  1. Requirements for the computer 
      Pentium III or Pentium IV processor-based personal computer 
     Windows 98/2000/XP or Windows NT4.0 
 
  2. Insert the CD-Rom which supplied with the instrument to the CD driver, and install the software step 
by step according to the instructions. The software  PV1785B_1788 and  PC1785B_1788  will be 
installed on your computer at the same time.