Scanner Installation
Installing the Scanner
2-7
With circuit power on and all OpScan iNSIGHT 70 system components 
unplugged, the voltage between the circuit neutral conductor and the isolated 
ground conductor must be less than 0.6 VAC.
Site Power Wiring Recommendations
Table 2-2 describes the conditions for power wiring at a scanner site.
Table 2-2  Power Wiring at a Scanner Site
Condition
Description
Dedicated 
Feeder
Run a separate dedicated feeder from the building main power panel to the 
branch power panel serving the OpScan iNSIGHT 70 system. The cabling 
should be of sufficient diameter to produce no less than 110V at the outlet 
when the scanner is connected and powered on. This feeder should be a 
minimum of 30 amp service to allow for future growth of the system.
Dedicated 
Isolated/ 
Insulated 
Ground
An isolated/dedicated ground is required. Run this ground conductor as an 
uninterrupted circuit from the power panel to the OpScan iNSIGHT 70 
system branch panel. The gauge of this conductor must be at least as large as 
that of the AC circuit conductors in all stages of the circuit.
Panel Branch 
Circuits
All branch circuits leaving the OpScan iNSIGHT 70 branch power panel 
should be at least 10 AWG copper wire. There should be no circuits from 
this power panel serving devices that are not part of the scanner system.
One branch circuit from this power panel should be dedicated at the scanner 
(this circuit will also service the computer and monitor), and there should be 
at least one additional dedicated circuit for each additional scanner.
Additional Utility 
Convenience 
Connectors
Provide additional convenience outlets in the vicinity of the OpScan 
iNSIGHT 70 scanner.  Connect these outlets (may be used for vacuums, 
paper joggers, and other electrical devices) to a branch panel other than the 
scanner branch panel to provide maximum noise isolation for the OpScan 
iNSIGHT 70 system components.
Scanner 
Electrical 
Connections
Dedicate one 15 amp (120 VAC) or 7.5 amp (220 VAC) circuit to each 
OpScan iNSIGHT 70 scanner at the site. All may be served from the same 
branch panel, provided its rated capacity is not exceeded.
Electrical 
Outlets
All 120V outlets for the OpScan iNSIGHT 70 scanner should be Hubble 
IG-5362 or equivalent (NEMA 5-20R). All 240V outlets should be Hubble 
IG2410A or equivalent. Outlets for the peripheral equipment should be of 
the type and rating specified in the hardware reference manuals for the 
equipment.
The quantity of outlets that must be provided for the peripheral equipment 
varies according to system configuration. (See Panel Branch Circuits above.) 
Outlets should be located within 6 feet (2 meters) of the working location of 
the equipment.