Model GFC7000E Instruction Manual 
Operating Instructions 
04584 Rev A1 
97 
 
or downloaded, edited and uploaded to the instrument in form of a string of plain-text lines 
through the communication ports. 
iDAS data are defined by the PARAMETER type and are stored through different triggering EVENTS 
in data CHANNELS, which relate triggering events to data parameters and define certain 
operational functions related to the recording and reporting of the data. 
6.12.1.1. iDAS Channels 
The key to the flexibility of the iDAS is its ability to store a large number of combinations of 
triggering events and data parameters in the form of data channels. Users may create up to 20 
data channels and each channel can contain one or more parameters. For each channel one 
triggering event is selected and up to 50 data parameters, which can be the same or different 
between channels. Each data channel has several properties that define the structure of the 
channel and allow the user to make operational decisions regarding the channel (Table 6-22). 
Table 6-22:   iDAS Data Channel Properties 
PROPERTY DESCRIPTION  DEFAULT SETTING 
RANGE 
NAME 
The name of the data channel. 
“NONE” 
Up to 6 letters and 
digits (more with 
APICOM, but only the 
first six are displayed 
on the front panel). 
TRIGGERIN
G EVENT 
The event that triggers the data channel to 
measure and store its data parameters. See 
APPENDIX A-5 for a list of available triggering 
events. 
ATIMER 
See Appendix A-5 For 
a complete list. 
NUMBER & 
PARAMETER 
LIST 
A User-configurable list of data types to be 
recorded in any given channel. See APPENDIX 
A-5 for a list of available parameters 
1 – DETMES  See Appendix A-5 For 
a complete list. 
STARTING 
DATE 
The starting date when a channel starts 
collecting data 
01-JAN-03 
Any actual date in the 
past or future. 
SAMPLE 
PERIOD 
The amount of time between each data point 
that is averaged into one mean reported every 
REPORT PERIOD. 
000:01:00 
000:00:01 to 
366:23:59 
(Days:Hours:Minutes) 
REPORT 
PERIOD 
The amount of time between each channel 
data point. 
000:01:00 
 
000:00:01 to 
366:23:59 
(Days:Hours:Minutes)
 
NUMBER OF 
RECORDS 
The number of reports that will be stored in 
the data file. Once the specified limit has been 
exceeded, the oldest data are over-written to 
make space for new data. 
100 
1 to 1 million, limited 
by available storage 
space. 
RS-232 
REPORT 
Enables the analyzer to automatically report 
channel values to the RS-232 ports. 
OFF 
OFF or ON 
CHANNEL 
ENABLED 
Enables or disables the channel. Provides a 
convenient means to temporarily disable a 
data channel. 
ON 
OFF or ON 
CAL HOLD 
OFF 
Disables sampling of data parameters while 
instrument is in calibration mode.  
When enabled here – there is also a length of 
the DAS HOLD OFF after calibration mode, 
which is set in the VARS menu  
OFF 
OFF or ON 
(Section 6.12.2.11.)