M400E Calibration Procedures 
M400E Ozone Analyzer Operator’s Manual 
9.4.1. SETUP  ACAL: PROGRAMMING AND AUTO CAL SEQUENCE 
To program the example sequence shown in Table 9-3, press: 
 
CONTINUE NEXT PAGE
With STARTING TIME 
SAMPLE                RANGE = 500.0 PPB               O3 
=XXX.X  
 
 SETUP X.X
 
 CFG   ACAL   DAS   RNGE   PASS   CLK   MORE   EXIT
 
 SETUP X.X
   SEQ 1) DISABLED
 
 
            
NEXT
  MODE   
 
 
       EXIT
 
 SETUP X.X
   SEQ 2) DISABLED
 
 
PREV  NEXT  
MODE
   
 
 
       EXIT
 
 SETUP X.X
   MODE: DISABLED
 
 
            
NEXT
     
 
 
         ENTR  EXIT
 
 SETUP X.X
    
MODE: ZERO 
 
PREV   
NEXT
     
 
 
         ENTR  EXIT
 
 SETUP X.X
   MODE: ZERO–HI
 
 
PREV   
NEXT
     
 
 
         
ENTR
  EXIT
 
 SETUP X.X
   SEQ 2) ZERO–HI,  1:00:00
 
 
PREV  NEXT  
MODE   
SET
 
 
 
      EXIT
 
 SETUP X.X
   TIMER ENABLE: ON
 
 
          
SET>
   EDIT 
 
 
     
       EXIT
 
 SETUP X.X
   STARTING DATE: 01–JAN–02
 
 
   <SET  SET>   
EDIT
   
 
 
       EXIT
 
 SETUP X.X
   
STARTING DATE: 01–JAN–02 
 
 
     
0
        4      SEP       
0
        3
   
    
 ENTR
     EXIT
 
Toggle keys to set  
Day, Month & Year: 
 
Format : DD-MON-Y
 SETUP X.X
   MODE: ZERO–LO
 
 
PREV   
NEXT
     
 
 
         
ENTR
  EXIT
 
 