M400E Ozone Analyzer Operator’s Manual 
38 
04315 Rev. C1 
3.7.4.2. Verifying the M400E Reporting Range Settings 
While it is possible to perform the following procedure with any range setting, we recommend that you perform 
this initial checkout using following reporting range settings: 
  Unit of Measure: PPB 
 Reporting 
Range: 
500 ppb  
 Mode Setting: SNGL 
These are the default setting for the M400E analyzer; however, it is a good idea to verify them before proceeding 
with the calibration procedure.  Use the same method as described in Section 3.6.2.1. 
 
3.7.4.3. Initial Zero/Span Calibration Procedure:  
Unlike other versions of the M400E, analyzers with the IZS option installed do not require the expected span gas 
concentration be set during initial start-up because no initial span calibration is performed.   
 
3.7.4.4. Initiate Daily Zero-Point Auto-Cal of M400E’s Monitoring Low Levels of O
3
  
To ensure that the analyzer maintains maximum performance levels when monitoring low levels of O
3
, the 
instrument’s AUTOCAL feature (only active on analyzers with the IZS option  installed) should be used to initiate 
a zero-point calibration once every day.   
The appropriate AUTOCAL sequence settings are: 
  Table 3-9:  AUTOCAL Settings for Daily Zero-Point Auto-Cal of M400E’s Monitoring Low Levels of O
3
 
MODE AND ATTRIBUTE 
VALUE 
COMMENT 
SEQUENCE 1 
Define sequence #1 
MODE ZERO 
Select zero calibration  mode 
TIMER ENABLE 
ON
1
 
Enable the timer 
STARTING DATE 
01 – JAN – 03
1
 
Start running sequence after January 
1, 2003 
STARTING TIME 
00:00
1
 
Start initial zero-point calibration at 
starts at midnight. 
DELTA DAYS 
1
1
 
Do Sequence #1 every day 
DELTA TIME 
00:00
1
 
Do Sequence #1 at the same time 
every day 
DURATION 15.0
1
 
Operate zero-cal valve for 15 min 
CALIBRATE ON 
The instrument will re-set the slope 
and offset values for the O
3
 
measurement calculation at the end of 
the auto-cal sequence 
1
 While most of the settings required for this sequence are the instrument’s default 
settings, and therefore do not need to be changed, they should be verified. 
If any of these settings do not match those shown, see the instructions in Section 9.4. 
To activate this daily zero point calibration: