Laser beta lasermike LS8000-3 Manuale Utente

Pagina di 221
LaserSpeed 8000-3 Instruction Handbook
 
RS-232/RS-422 Communication 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 125 of 221 
Revision A (Sep 2007) 
Method 2 
1)  Start HyperTerminal (or any other terminal program) and set to 19200, 
7, N, 2 (or the baud rate you want to use later).  HyperTerminal is 
installed with Windows 2000/XP, and can be found in  Start 
Menu Programs Accessories Communications 
2)  Cycle the gauge power. 
3)  Hit the Enter key 3 times, or transmit three 
Carriage Returns (hex 0x0d). 
4)  Use the "Z" command to verify gauge is 
communicating. 
5)  Use the "J" command to set the RS-232 
baud rate or the "@" command to set the 
RS-422 baud rate. 
6)  Reconfigure HyperTerminal to the new 
baud rate if different from that in step 1. 
Communication Protocol 
This section lists the commands used to configure the LS8000-3, and to read 
measurement data.  Throughout this section, <CR> is used to denote a 
carriage return, and <LF> is used to indicate a linefeed. 
Notation 
Description 
<CR> 
Carriage Return character 
ASCII Value: 13 
<LF> 
Linefeed character 
ASCII Value: 10 
<N> 
Integer Value (variable length) 
Indicates an integer value with length from 1 to 9 characters.  When sending 
a command to set a configuration setting, the transmitted value can contain 
leading zeros, but they are not required.  When reading a configuration 
setting, the LS8000-3 will not transmit leading zeros in its response. 
Example Notation: L<N><CR> 
Example Command: L100000<CR> 
<IP> 
IP Address 
Example Notation: *IPADDR=<IP><CR> 
Example Command: *IPADDR=192.168.10.246