Ethernet Communication 
The LS8000-3 has an optional Ethernet port that can be used to collect data 
and configure settings.  Measurement data and configuration settings are 
available over both the TCP and UDP protocols.  For ease of use, the LS8000-
3 can act as a DHCP client, automatically obtaining a network IP address from 
a DHCP server.  Also, it supports host name advertisement, so that the gauge 
can be addressed on the network without actually knowing the IP address. 
Supported Protocols 
In general, TCP should be used for configuration settings and UDP for real-
time measurement data.  All data and configuration commands are available 
over both protocols, but UDP has much lower overhead, so it is better suited 
for continuous, high-speed, data output.  TCP is a slower protocol that 
guarantees delivery of each packet, so it is better suited for configuring 
settings.  TCP and UDP can be used concurrently. 
COM LED 
The COM light on top of the LS8000-3 can be configured to indicate Ethernet 
activity by setting the COM LED Function ($C command) setting to 1.  This is a 
blue-orange LED that can have the following states: 
Ethernet State 
COM LED 
No Connection 
Off 
Connection OK 
Orange 
Activity Detected 
Blue 
Collision White 
Configuration Settings 
This table lists the configuration settings relevant to Ethernet communications. 
Setting 
Configuration 
Command 
Read/Write 
Ethernet Enabled 
*ETHERNET 
Read Only 
Ethernet Hardware ID 
*MACID 
Read Only 
DHCP Enable/Disable 
*DHCP 
Read/Write 
Host Name 
*NAME 
Read/Write 
IP Address 
*IPADDR 
Read/Write 
Default Gateway Address 
*IPDEFGW 
Read/Write