Laser beta lasermike LS8000-3 Manuale Utente

Pagina di 221
LaserSpeed 8000-3 Instruction Handbook
 
Appendix E: Troubleshooting Guide 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 205 of 221 
Revision A (Sep 2007) 
Appendix E: Troubleshooting Guide 
Communications 
 
Can't connect to the gauge 
through the RS-232/RS-422 serial 
port. 
 
--- or --- 
 
LaserTrak won't detect the gauge. 
 
--- or --- 
 
Can't communicate with terminal 
program (such as HyperTerminal) 
 
 
 
 
 
 
 
Wiring or wrong Baud Rate 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Computer Problem 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1.  Cycle the gauge power and check the 
baud rate and framing as described 
under 
Baud Rate Indication
2.  Verify that you are connected to the 
proper pins.  The 9-pin DSUB on the 
back of the gauge is not an RS-232 
port. 
3.  Verify that the serial cable you are 
using is not a null modem cable.  A 
straight-through cable is required. 
4.  Verify that the breakout is wired 
correctly: 
DB25 Pin 1   DB9 Pin 2 
DB25 Pin 2   DB9 Pin 3 
DB25 Pin 19   DB9 Pin 5 
 
 
Verify that the serial port and cable are 
useable by the software. 
1.  Connect to the serial port with 
HyperTerminal 
2.  Make sure the "Echo typed 
characters locally" checkbox is NOT 
checked.  Set the Hyperterminal 
settings to match the following: 
 
 
3.  Do a loop-back test by disconnecting 
the serial cable from the LaserSpeed 
breakout DB9 and short pins 2 and 3 
of the cable together with a paperclip 
or screwdriver. 
4.  You should be able to type on the 
keyboard and see everything that you