Laser beta lasermike LS8000-3 Manuale Utente

Pagina di 221
LaserSpeed 8000-3 Instruction Handbook
 
Installing the System 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 24 of 221 
Revision A (Sep 2007) 
Connecting Power to the Gauge 
The gauge requires a 20–28 VDC (nominal 24 VDC) supply to be connected to 
Pin 24 and/or Pin 25 with the corresponding ground to Pin 12 and/or Pin 13 of 
the 25-pin D-sub connector. The supply should be able to drive at least 2 
Amps. For detailed information on supplying power to the gauge, see 
Appendix D
Notes:   Do not attach wires to the gauge while power is applied. Make all 
wiring connections with power turned off. The gauge contains internal 
fusing (3A, 125 V) on the 24 VDC supply line. This fuse is not user 
serviceable.  
The LS8000-3 is classified as a kit, according to CDRH regulations. It is your 
responsibility to install a key switch to control the operation of the device. The 
key switch should be installed as shown in the following figure. It is in line with 
the 24-volt supply line that will service to power the device. The recommended 
key switch can be found in the accessory kit. The key switch should be 
mounted in a convenient location that does not require exposure to the laser 
beams. After installation, label the panel as to which position corresponds to 
“Laser On” and “Laser Off.” 
The following figure shows the correct installation of the recommended switch 
(C&K Components part number YM06132C205NQ). If another type of key 
switch is used, it should be verified that the key is removable only in the “Laser 
Off” position. 
For more information on laser safety issues, refer to the 
Safety 
section at the 
beginning of this manual. 
Warning:   A key switch must be installed and labelled correctly to ensure 
proper protection of personnel working with the laser.  
Panel Cutout
for Keyswitch
0.42in
[10.7mm]
0.50in
[12.7mm]
Laser Off
Laser On