LaserSpeed 8000-3 Instruction Handbook
 
Interfacing with the LS8000-3 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 59 of 221 
Revision A (Sep 2007) 
0
2
4
6
8
10
12
14
16
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
Quality Factor
A
nal
og O
ut
put
 /
 V
ol
ts
Analog Output vs Quality Factor
Voltage = Measured Analog Output Voltage 
 
Note that when the output voltage is at either limit, the exact speed cannot be 
determined.  If the exact speed is always required, be sure to set the Full 
Scale Velocity setting to a value higher than what can normally be 
encountered. 
The analog port can also be configured to 
output the QF on a 0–1 V scale by setting 
the Full Scale Velocity to zero using a serial 
command or LaserTrak software.  In this 
case, the analog output reads 0V when 
the QF is 0 and 1V when the QF is 15. 
When the Analog Output is configured to 
output Quality Factor, the Quality Factor 
can be calculated with the following 
equation: 
15
×
Voltage
tor
QualityFac
 
When connecting to the Analog Output Voltage, the Analog Output Voltage 
and Analog Output Ground should be on a separate twisted pair inside of a 
shielded cable.  Connect a 0.1
µF capacitor across the load end of the cable 
(the end opposite the gauge) to reduce noise caused by other signals in the 
cable.  See the following wiring example: 
 
Gauge DB9
0.1 F
µ
1000
minimum
Ω
Analog Output Ground (Pin 7)
Analog Output+ (Pin 6)
Analog Output Cable
Shielded, Twisted Pair
Cable Jacket
Cable Shield
Terminate Shield/Drain Wire to backshell of DB9 connector
Terminate Shield/Drain Wire to
Chassis Ground
 
Analog Output Wiring Example 
 
Measurement Synchronization Input (Pins 8-9) 
In many AGC applications, particularly mass flow control, it is desirable to 
synchronize the sampling periods of all gauges in the system.  This is done by 
connecting an LS8000-3 I/O Module to each LS8000-3 in the system (for