Invacare Electric Portable Patient Lift RPA600-1E Manuale Utente

Pagina di 40
SECTION 8—ACCESSORIES
37
3. Press the CAL button located on the PC board. The CAL switch is not labeled but is 
the only button on the PC board located in the lower right corner. Once the CAL 
button is pushed ʺCAL1ʺ will be seen in the display window.
NOTE: CAL1 selects the calibration mode using 50 pounds of calibrated weight. CAL2 selects the 
calibration mode for use with 200 pounds of calibrated weight. Pressing the UNITS key toggles 
between CAL1 and CAL2.
4. When the desired calibration mode is displayed, press the LOCK/UNLOCK key. 
NOTE: The display window will now show UnLd.
5. Ensure that there is no load on the scale and press the ZERO key.
6. The display window will show a dashed line (‐‐‐‐‐‐) scrolling across and then the word 
LOAD will appear in the display window.
7. Perform one of the following:
A. FOR CAL1 OPTION ‐ support 50 pounds of calibrated weight from the scale and 
press the ZERO key.
B. FOR CAL2 OPTION ‐ support 200 pounds of calibrated weight from the scale and 
press the ZERO key.
8. The display window will show a dashed line (‐‐‐‐‐‐) scrolling across and then the word 
DONE will appear in the display window.
9. Remove the weight from the scale and press the LOCK/UNLOCK key.
NOTE: The unit will now be in a temporary test mode and will not lock the weight display. This 
will allow weight to be loaded and unloaded to check the calibration. 
10. Turn the unit off by pressing the ON/OFF key. 
11. Turn the unit on by pressing the ON/OFF key.
NOTE: Scale will now be in normal operation.
Troubleshooting
Display Codes
CALIBRATION REQUIRED ‐ Indicates improper stored calibration data, calibration is 
necessary.
OVER CAPACITY ‐ Indicates a weight exceeding the capacity has been loaded on the scale.
SYMPTON
PROBABLE CAUSE
SOLUTIONS
Unit does NOT work properly.
Battery failure.
Check battery. Replace if necessary.
Battery has been replaced and unit 
still does NOT work properly.
Contact Invacare for Service at 1-800-333-
6900
CAL
OCAP