RM500SL User’s Guide Version 2.8  
Page 33 
7.8 
HIT calibration facts 
FastFacts 7.8: HIT calibration facts 
HIT microphone calibration establishes a correction curve for the uncalibrated HIT 
reference microphone by comparison with the factory calibrated coupler 
microphone. It may be performed as often as desired but will be invalidated at 1 
am the following day or at 1 am the following Monday, depending on the choice of 
Daily or Weekly calibration interval. After the calibration has expired, you will be 
prompted to calibrate whenever you attempt a test that uses the HIT microphones. 
Note that there are two styles of HIT reference microphones. The newer style is 
larger but responds beyond 16 kHz and has a lower noise floor. The two have 
different correction curves, as shown below, but will produce the same test results, 
once calibrated. Your calibration curve should be similar to one of these and 
should remain stable over time. Irregularities in the calibration curve may be the 
result of positioning errors. See 
7.9: Calibration of HIT reference microphone.