RM500SL User’s Guide Version 2.8  
Page 42 
8.7 
ANSI input-output curves 
FastFacts 8.7: ANSI input-output curves 
An input/output curve shows the output level of a hearing instrument as a function 
of input level and frequency. It is a steady-state pure-tone curve and does not 
reflect performance with complex signals such as speech. 
You may run the tests at any setting of the instrument controls but ANSI S3.22 
requires that certain settings be used. See 
8.2: ANSI S3.22-1996 facts
 and 
8.3: 
ANSI S3.22-2003 facts
. The [Input/Output] test allows you to generate up to 4 
separate input/output curves at each of 5 individual frequencies (250, 500, 1000, 
2000, 4000 Hz) as provided in ANSI S3.22. 
1)  Press <Tests>, then highlight and <PICK> [Input/Output]. 
2)  Refer to 
7.11: Coupling the hearing instrument
7.12: Positioning the hearing 
instrument
 and 
2.5: Battery pill use and care
 to prepare the instrument for 
testing. 
3)  Turn the instrument ON; set AGC controls or programming parameters as in 
8.4: ANSI 1996 Linear and AGC tests
 or 
8.5: ANSI 2003 Linear and AGC tests
  
4)  Highlight and <PICK> [Curve 1]. Test setup instructions will be displayed. 
5)  Press <Continue> to generate a set of 5 input/output curves. 
6)  To measure another set of 5 input/output curves, highlight and <PICK> another 
[Curve #]. 
7)  To summon a cursor to determine curve values, position the mouse pointer on 
a graph area and click the left mouse button. Click again to turn it off. 
NOTE: Noise in the test area may cause erroneous results at low input levels 
(particularly at 250 and 500 Hz). Ensure that the test chamber is well sealed and 
ambient noise is low.