9.5 
Spectral analysis in Multicurve 
FastFacts 9.5: Spectral analysis in Multicurve 
9.6 
Battery drain test 
FastFacts 9.6: Battery drain test 
Selecting a stimulus level of 0 dB (off) in [Multicurve] presents a spectral analysis 
of the coupler microphone signal (Scale = dB SPL) or the difference between the 
coupler and reference microphone spectra (Scale = dB Gain). 
If the stimulus type is [Pink noise], the analysis is performed in 1/12
th
 octave bands 
and displayed in real time. Pressing <Continue> captures the displayed spectrum. 
If the stimulus type is [Swept], a swept filter analysis is performed in 1/12
th
 octave 
bands and displayed at 1/3
rd
 octave frequencies. Pressing <Continue> causes the 
swept filter analysis to be performed in 1/12
th
 octave bands and displayed at 1/12
th
 
octave frequencies. 
The battery drain test automatically measures the battery current when there is no 
input signal (quiescent), when the input signal is a 1 kHz tone at 65 dB SPL and 
the average of the currents when the input signal is 90 dB SPL at the HFA or SPA 
frequencies (Avg @ 90 dB). Battery life is estimated assuming 80% quiescent and 
20% Avg @ 90 dB use. Optional battery pills are required
With the hearing instrument set as normally used: 
1)  Install and connect the proper battery pill. See 
2.5: Battery pill use and care
 
2)  Highlight and <PICK> Battery drain from Hearing Instrument Tests list. 
3)  Highlight and <PICK> [Battery type] and select from drop-down list. 
4)  Highlight and 
<PICK> [Hours 
per day] and 
select from drop-
down list. 
5)  Highlight and 
<PICK> [Start 
test]