Honeywell Water System 4973 Series Manuale Utente

Pagina di 26
Specifications 
Installation and Maintenance Manual 
10/07 
 
 
2. Specifications 
2.1  Specifications for 04973 Series 
Parameter Description 
Cell Constant 
04973 Series: 0.01, 0.1, 1.0 and 10 as specified 
Electrode Material 
0.01 and 0.1 constant, titanium 
1.0 and 10 constant, high density graphite with 
Teflon sheath 
Maximum Pressure Limit 
1724 kPa (250 psig) at rated temperature 
Maximum Continuous Temperature Limit 
140°C (284°F) at rated pressure 
For PVC wire: 
105°C (221°F) 
Materials of Construction 
Cell Body: PES (polyethersulfone) 
Electrodes: see ‘Electrode Material’ 
Cable Options 
Leadwire: PVC insulated 18 gage cable 
20 and 50 feet lengths available 
Quick Disconnect Option 
Universal Head (Aluminum) 
Approvals 
Manufactured to comply with ASME boiler and 
pressure vessel code Section III, Div.1, UG-101 
CRN #0F11607.5C 
Insertion 
3/4” NPT male, Schedule 40 and 80 pipes 
Insertion Depth 
3-1/2” (89 mm) for 1, 10 and 0.01 cell constants 
from solution end of 3/4" MNPT 
2-1/2” (64mm) for 0.1 cell constant 
 
2.2  Specifications for 055919 Flow Chamber 
Parameter Description 
Maximum Flow 
2 gpm @ 40psig and atmospheric discharge 
Maximum Pressure 
200 psig @ 25°C 
Maximum Temperature 
140°C (284°F) at atmospheric pressure 
Dimensions 
See Figure 3-1 
Materials of Construction 
Polyethersulfone (PES)