Hitachi Microscope & Magnifier S-4800 Manuale Utente

Pagina di 294
 
3 - 48
 
NOTICE
 
When the specimen height is set higher than the standard position, actual specimen surface 
position will be changed by tilting. The allowable range of Z and Tilt are therefore recalculated 
after the tilting angle is specified for movement. It may cause somewhat different allowable 
range value than that has been indicated before specifying tilting angle.
   
   
(2) Modes of operation 
Two modes of operation, "Set values and Go" and "Continuous driving" are possible. 
When the Continuous mode is selected, the stage moves following the scrollbar operation for 
each axis. It is recommended to press or click the arrow buttons of the scrollbars, as the 
mechanical motion of the stage is relatively slow compared to the scrollbar operation.   
The step of motion is 0.1mm (Z), 0.5
°
 (Tilt) by arrow buttons and 1mm (Z), 1
°
 (Tilt) by clicking 
on scrollbar (side of slider). 
It is advisable to use mouse operation on the scanning image for continuous operation of Z or 
Tilt axes. 
 
When the Continuous mode is off (unchecked), the stage begins its motion after clicking the 
Go button. In this mode of operation, select Z and Tilt movement with scrollbar s and then 
click the Go button. 
Blue numerals on scrollbars indicate the movable range of each axis. These values are 
adjusted according to the specimen size and present position of other axes. 
 
NOTICE: Release stage lock to enable Z and Tilt operation. 
 
 
 
CAUTION
 
 
The specimen size and height, and detectors being used must be set correctly to 
allow motion of each axis in the proper ranges. If this is not done, the specimen may 
strike the objective lens and cause damage to the specimen and objective lens