Hitachi Microscope & Magnifier S-4800 Manuale Utente

Pagina di 294
3
-
93
(d) Save condition 
Click the Save button. The Condition Save dialog window will open. 
Input comment, specify a Condition name and then, click Save button.   
 
 
(2) Loading a Set of Operating Conditions 
Use the following steps to load a set of operating conditions. 
 
(a) Open the Setup dialog window   
Select Condition Load from the Operate menu, or click 
the Setup button in the tool button section of the 
operation panel and select Op.Cond tab.   
Select Load Memorized Condition on it. 
(b) Select a condition name 
Memorized condition names are listed. Select a name by 
clicking it. 
Use sample picture and comment for reference of 
selection. 
(c) Select items to be loaded 
Check items ( Vacc to Contrast) you wish to load and 
reproduce. Stage position is effective only for Type2 
instrument. 
(d) Reproduce column condition 
Click the Load button. Memorized data are retrieved and 
the electron optical column is set.   
 
If the Stigma / Alignment is included, a message will 
appear for confirmation. These are possibly changed if 
the loaded condition was saved long before. In such 
cases, it may be better not to reproduce Stigma and 
Alignment data. The recent data is memorized 
automatically by another condition memory function (the 
most recent condition for each of Vacc 
 Condenser lens setting is memorized and 
reproduced automatically)   
     
(3) Deleting an Operating Conditions   
Select a condition name as same manner as above, and click the Delete button.