Specifications
Automatic Scope Measurements
9
93
Power
Power Factor ..................... ratio between Watts and VA
Range ......................................................0.00 to 1.00
Watt................................. RMS reading of multiplication
corresponding samples of input A (volts)
and Input B (amperes)
Full Scale Reading .................................... 999 counts
VA ............................................................ Vrms x Arms
Full Scale Reading .................................... 999 counts
VA Reactive ................................................
((VA)
2
-W
2
)
Full Scale Reading .................................... 999 counts
Phase
Range.......................................... -180 to +180 degrees
Resolution....................................................... 1 degree
Accuracy
0.1 Hz to 1 MHz ........................................
±
1 degrees
1 MHz to 10 MHz ......................................
±
3 degrees
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°
C or 
°
F) ............................... -40.0 to +100.0 
°
-100 to +250 
°
-100 to +500 
°
-100 to +1000 
°
-100 to + 2500 
°
Probe Sensitivity ..........................1 mV/
°
C and 1 mV/
°
F
Decibel (dB)
dBV..............................................dB relative to one volt
dBm ................. dB relative to one mW in 50 
Ω
 or 600 
Ω
dB on ........................................VDC, VAC, or VAC+DC
Accuracy.........................same as VDC, VAC, VAC+DC