Fluke ScopeMeter 196 Manuale Utente
Specifications
Automatic Scope Measurements
9
93
Power
Power Factor ..................... ratio between Watts and VA
Range ......................................................0.00 to 1.00
Watt................................. RMS reading of multiplication
corresponding samples of input A (volts)
and Input B (amperes)
Full Scale Reading .................................... 999 counts
VA ............................................................ Vrms x Arms
Full Scale Reading .................................... 999 counts
VA Reactive ................................................
√
((VA)
2
-W
2
)
Full Scale Reading .................................... 999 counts
Phase
Range.......................................... -180 to +180 degrees
Resolution....................................................... 1 degree
Accuracy
0.1 Hz to 1 MHz ........................................
±
1 degrees
1 MHz to 10 MHz ......................................
±
3 degrees
Temperature (TEMP)
With Optional Temperature Probe
Ranges (
°
C or
°
F) ............................... -40.0 to +100.0
°
-100 to +250
°
-100 to +500
°
-100 to +1000
°
-100 to + 2500
°
Probe Sensitivity ..........................1 mV/
°
C and 1 mV/
°
F
Decibel (dB)
dBV..............................................dB relative to one volt
dBm ................. dB relative to one mW in 50
Ω
or 600
Ω
dB on ........................................VDC, VAC, or VAC+DC
Accuracy.........................same as VDC, VAC, VAC+DC