The WLK 10 is an air-cooled refrigeration
aggregate designed to facilitate anode
heat dissipation of the X-ray tube when
operating at high ambient temperatures
aggregate designed to facilitate anode
heat dissipation of the X-ray tube when
operating at high ambient temperatures
(up to 40° C) and can also be used when,
for economical reasons cooling of the
X-ray tube from the water supply is not
feasible.
for economical reasons cooling of the
X-ray tube from the water supply is not
feasible.
Technical Data
Dimensions and weight
height
485 mm / 19.1”
width
352 mm / 13.9”
length
630 mm / 24.8”
weight
approx. 65 kg (143 lbs) without water
capacity
7 litres
Characteristics
cooling power
1200 W at 25°C ambient temperature (forward line temp. 15° C)
1050 W at 32°C ambient temperature
900 W at 40°C ambient temperature
waterflow rate
> 4.4 l/min at 4.0 bar
power connection
230 V, 50/60 Hz
max. mains current
7.5 A
noise level
< 70 dB(A) (in a distance of 1 m / 3.94 ”)
Environmental conditions
operating temperature
- 5° C ... + 40° C (with antifreeze filling)
storage temperature
- 25° C ... + 70° C (with antifreeze filling)
humidity
20 % ... 80 %
Coolant circuit
cooling agent
R 134A (CFC-free)
Settings
pressure
< 6.2 bar at 50 Hz
forward line
18° C
max. temperature
35° C
Connecting cable
3 x 1.5 mm
2
;
Grounding plug; 10 m (32 ft) length
Application