Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
Operational Checks
11-7
11
Battery Capacity Test (CT)
To perform a Battery Capacity Test:
1. Turn the HeartStart XLT off.
2. Place a "Test in Progress" label on the HeartStart XLT to indicate to oth-
ers that it may not be used.
3. Insert a charged battery.
4. If an AC power module is connected, unplug the power module from the 
HeartStart XLT. While pressing 
, press On to start the test.
5. Allow the test to proceed to completion. The test takes approximately 
three hours and is complete when test results print out and the device 
turns itself off. 
6. Review the test results and take the appropriate action, as follows:
Table 11-1  Battery Capacity Test Results
If 
Then...
Elapsed Time > 2.5 hours
              and
Low Battery Time > 10 minutes
1.
The battery passed the test.
2.
Record "pass CT" and the date on 
the bottom of the battery.
3.
Recharge the battery before use.
Elapsed Time < 2.5 hours
               or
Low Battery Time < 10 minutes
1.
The battery failed the test.
2.
Record "fail CT" and the date on 
the bottom of the battery.
3.
Discard the battery appropriately.
Mark
Event