Specifications
13-9
13
Environmental
Temperature: 0
o
 to 50
o
C operating, -20
o
 to 70
o
C storage.
z
Thermal paper may darken above 55
o
C.
z
Charging the battery at temperatures above 35
o
C may degrade battery 
life.
z
Storing the battery for extended periods at temperatures above 40
o
will reduce battery capacity and degrade battery life.
z
Operating and storage specifications for electrodes may vary. Refer to 
the manufacturer’s specifications for details.
Humidity: 
Up to 95% Relative Humidity
z
Printer paper may jam if paper is wet.
z
Thermal printer may be damaged if wet paper is allowed to dry while 
in contact with printer elements.
Altitude: 
Operating: up to 15,000 ft.
Shipping: up to 20,000 ft.
Shock (drops onto concrete): Unit survives 39" (1 m) drops onto all surfaces 
(faces, corners, and edges) enclosed in carrying case. Exposed handle surfaces 
survive 24" drops.
Vibration: Mil Std 810E 514.4 Category 6 Helicopter, General Storage, 
UH60.
Water Resistance: IPX4, splash proof per IEC 60529, with Data Card door 
closed.
EMC: Meets EN 60601-1-2.  This ISM device complies with Canadian 
ICES-001.
Other Considerations: Equipment not suitable for use in the presence of a 
flammable anesthetic mixture with air, oxygen, or nitrous oxide.
Mode of Operation: Continuous.