Safety Considerations
13-20
Specifications & Safety
WARNING
Do not allow multifunction defib electrode pads to touch each other or to touch 
other ECG monitoring electrodes, lead wires, dressings, etc. Contact with metal 
objects may cause electrical arcing and patient skin burns during defibrillation and 
may divert current away from the heart.
WARNING
During defibrillation, air pockets between the skin and multifunction defib electrode 
pads may cause patient skin burns. To help prevent air pockets, make sure the pads 
completely adhere to the skin. Do not use dried out pads; do not open pads package 
until just prior to use.
WARNING
Never touch the patient or any equipment connected to the patient (including the 
bed or gurney) during defibrillation.
WARNING
Do not operate the HeartStart XLT in standing water. When using the HeartStart XLT 
in wet environments, make sure the Data Card door is securely shut. 
WARNING
Do not immerse, or pour fluids on, any portion of the HeartStart XLT. 
WARNING
Do not use the HeartStart XLT in a flammable or oxygen-rich atmosphere. This can 
cause an explosion hazard.