Troubleshooting
3-18
Defibrillating in AED Mode
Troubleshooting
When the HeartStart XLT detects a problem, it provides display and/or voice 
prompts to guide you to resolution. The table below lists the prompts you may 
encounter in AED Mode, the cause, and the suggested corrective action. 
Prompts related to the battery and Data Card are discussed in Chapter 12.
Table 3-1  AED Mode Prompts
Prompt
Possible Cause
Corrective Action
Pads Off
The multifunction defib electrode 
pads are not properly applied to the 
patient.
Check that the pads are applied to 
the patient, as directed on the pads’ 
package. Replace the pads if the 
prompt continues.
Pads Cable 
Off
The pads cable is not connected to 
the defibrillator.
Check that the defibrillation pads 
connector is locked in place.
Artifact 
Detected
z
Patient motion interferes with 
analysis.
z
Electrical sources are causing 
interference.
z
Attempt to eliminate patient 
motion. Avoid analyzing 
during transport or while 
performing CPR.
z
Move hand-held communication 
devices or other suspected 
devices away from the defibril-
lator, when possible.
Shock Can-
celled
Shock key not pressed within 30 
seconds.
Press within 30 seconds of prompt.
No Shock 
Delivered
Pads are not properly connected to 
the patient.
Check pads connection.
Key 
Inactive
z
The key pressed only functions 
in Manual Mode.
z
The key pressed does not func-
tion during analysis or charging.
z
The key pressed does not func-
tion while in a pause state.
z
Access manual mode prior to 
pressing the key.
z
Wait for analysis or charging to 
complete prior to pressing the 
key.
z
Press 
 
prior to pressing 
the key.
 