Preparing for Synchronized Cardioversion
7-2
Performing Synchronized Cardioversion
Preparing for Synchronized Cardioversion
In preparation for synchronized cardioversion:
1. Apply multifunction defib electrode pads as directed on the package. Use 
either the anterior-anterior or anterior-posterior placement, as appropriate.
2. Connect the pads to the patient cable. (See Figure 6-2.)
3. Apply monitoring electrodes, if desired. (See “Applying Monitoring Elec-
4. If needed, insert a Data Card (as described in “Using a Data Card” on 2-
5. Press On.
7. Use 
 to select the best lead that displays a large QRS complex. (See 
Delivering a Synchronized Shock
To perform synchronized cardioversion:
1. Press 
 to enable Sync Mode. The green LED next to 
 
lights up and the message 
6<1& appears on the display.
2. Use the gain control, 
, to adjust the ECG size so that the marker dot 
appears only once with each R-wave.
3. If necessary, change the energy setting (the default setting is 150 joules), 
by pressing 
 repeatedly until the desired energy level is displayed. 
The choices are 5, 10, 25, 50, 70, 100, 150, or 200 joules.
4. Press 
. Wait until the current charge has reached the energy level 
selected and you hear a continuous charge done tone. 
/($'
6(/(&7
Sync
Sync
(1(5*<
6(/(&7
&+$5*(