HP Cell Phone Accessories E6380-90019 Manuale Utente

Pagina di 228
174
Chapter 6 
Configuration - Screens and Control Fields
Ref Select
Ref Select
Ref Select
This field selects the timebase reference signal to use for generating 
and analyzing signals.
Auto
 automatically selects an external reference if a signal of 
sufficient level is detected at the EXT REF IN. 
Internal
 uses the Test Set’s internal timebase and ignores signals 
connected to the EXT REF IN connector.
External
 requires a valid reference signal to be connected to the 
EXT REF IN connector.
Operating Considerations
The front-panel REFERNCE indicators show which source is selected 
and whether or not the timebase is phase locked. The Test Set will not 
operate correctly if it is unlocked.
Locking to an external reference may take several seconds. The lower 
the frequency the longer it may take to lock to the signal.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Main Menu