Example Program for CR10X 
;{CR10X} 

*Table 1 Program 
 01: 1 
 Execution 
Interval 
(seconds) 
 
1:  AC Half Bridge (P5) 
 1:  1 
 Reps 
  2:  25 
  2500 mV 60 Hz Rejection Range 
 3:  1 
 SE 
Channel 
  4:  1 
  Excite all reps w/Exchan 1 
 5:  2500 
 mV 
Excitation 
  6:  1 
  Loc [ V_Vx      ] 
 7:  1.0 
 Mult 
 8:  0.0 
 Offset 
 
2:  Z=1/X (P42) 
  1:  1 
  X Loc [ V_Vx      ] 
  2:  2 
  Z Loc [ Vx_V      ] 
 
3:  Z=X+F (P34) 
  1:  2 
  X Loc [ Vx_V      ] 
 2:  -1 
 F 
  3:  3 
  Z Loc [ Vx_V_1    ] 
 
4:  Z=X*F (P37) 
  1:  3 
  X Loc [ Vx_V_1    ] 
 2:  24900   F 
  3:  4 
  Z Loc [ Rtherm    ] 
 
5:  Z=LN(X) (P40) 
  1:  4  
  X Loc [ Rtherm    ] 
  2:  5 
  Z Loc [ lnRt      ] 
 
6:  Z=X*F (P37) 
  1:  5 
  X Loc [ lnRt      ] 
 2:   
.001 
 F 
  3:  6 
  Z Loc [ Scal_lnRt ] 
 
7:  Polynomial (P55) 
 1:  1 
 Reps 
  2:  6 
  X Loc [ Scal_lnRt ] 
  3:  7 
  F(X) Loc [ 1_Tk      ] 
 4:  .001129   C0 
 5:  .234108   C1 
 6:  0.0 
 C2 
 7:  87.7547   C3 
 8:  0.0 
 C4 
 9:  0.0 
 C5 
 
8:  Z=1/X (P42) 
  1:  7 
  X Loc [ 1_Tk      ] 
  2:  8 
  Z Loc [ Tk        ]