Standalone/Self Calibration
Every MultiSync MD Series display 
includes a standalone or self-calibration 
feature. This feature uses an external 
sensor (NEC MDSVSENSOR2 or X-Rite 
i1 Display2) plugged into a USB port 
on the display to perform a DICOM 
calibration without any software. 
This calibration option is useful for situations where a 
computer is not available to run the GammaCompMD 
QA software. It does not provide any information on the 
parameters measured during calibration, so standalone 
calibration may not be ideal for ongoing quality assurance.
Quick QA
MultiSync MD211C2 and MD211C3 models include a Quick 
screen QA feature, available at the push of a button. This 
feature uses the built-in front calibration sensor on the 
display and is designed to allow the user to perform a 
simple visual test, while providing a measurement of the 
luminance and DICOM GSDF error percentage. 
This QA test is ideal for times when a simple visual test is 
needed and there is a method for documenting the results, 
as the results of this test are not stored with the display. 