Fisher & Paykel MR810 Manuale Utente

Pagina di 43
 
 
MR810 Technical Manual Revision C  REF 185042601
 
31 
 
7
7 E
LECTRICAL 
S
AFETY 
T
ESTS
 
After the MR810 has been serviced, perform the following electrical safety tests as listed below or in 
accordance with relevant country regulations. 
7.1 P
ROTECTIVE 
E
ARTH 
R
ESISTANCE
 
Use an ohmmeter and measure the resistance from the earth pin on the mains plug to the heater plate.  To 
ensure good contact with the anodised heater plate an alligator clip is recommended.  Connect at the front of 
the heater plate where the anodising has been removed (see Figure 7.1).  The resistance should be 
 0.2
 Ω.  
 
7.2 E
ARTH 
L
EAKAGE 
C
URRENT
 
Measure earth leakage with the earth probe of the tester on the heater plate (see Figure 7.1).  Earth leakage 
current is measured at normal mains operating voltage and frequency; it should not exceed 0.5mA 
maximum. 
7.3 I
NSULATION 
R
ESISTANCE
 
Use a 500 VDC insulation resistance tester to measure the resistance from the phase pin on the mains plug 
to the heater plate (see Figure 7.1).  Insulation resistance should be 
 10 M
Ω.
 
 
 
Figure 7.1: Location of test point on heater plate