Fisher & Paykel MR810 Manuale Utente

Pagina di 43
 
40 
MR810 Technical Manual Rev C  REF 185042601
 
 
 
D 
Appendix D:  E
LECTROMAGNETIC 
C
OMPATIBILITY
 
D.1 C
OMPLIANCE AND 
W
ARNINGS
 
The MR810 with RT308 breathing circuit has been tested to and found compliant with the requirements of 
IEC 60601-1-2: 2001. 
WARNING: 
The use of other accessories other than those specified may result in increased 
emissions or decreased immunity of the equipment. 
Verify correct operation if the MR810 is to be used adjacent to other equipment. 
D.2 E
LECTROMAGNETIC 
E
MISSIONS
 
Guidance and manufacturer’s declaration – electromagnetic emissions 
The MR810 is intended for use in the electromagnetic environment specified below.  The customer or the user 
of the MR810 should assure that it is used in such an environment. 
Emissions Test 
Compliance 
Electromagnetic environment - guidance 
RF emissions 
CISPR 11  
Group 1 
The MR810 uses RF energy only for its internal 
function.  Therefore, its RF emissions are very low 
and not likely to cause any interference in nearby 
electronic equipment. 
RF emissions 
Class A 
Harmonic emissions 
IEC 61000-3-2 
Class A 
Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3 
Complies 
The MR810 is suitable for use in all establishments 
other than domestic and those directly connected to 
the public low-voltage power supply network that 
supplies buildings used for domestic purposes.