110  Stratos LV/LV-T Technical Manual   
2.18.6 General Statistical Information 
•  The Stratos CRT-P’s statistics modes are always in 
operation and cannot be selected OFF.  
•  The counters within the statistic features do not operate 
when a magnet is applied to the CRT-Ps 
•  The counters within the statistic features are reset each 
time the Stratos CRT-Ps are permanently programmed. 
•  Event counters are displayed as bar charts showing the 
event totals expressed as a percentage. 
•  Histograms count how often events occur in different 
time or rate intervals (for example, how many events 
occurred in a 160 to 169 ppm range). 
•  Trends represent a certain number of events at a fixed 
point in time (e.g., rates).  The trends are plotted as 
points that are joined together by a curve.  In the 
Stratos 
CRT-Ps, the trends of the three different 
channels are identified with different colors. 
 
N
OTE
When a magnet is applied, Stratos CRT-Ps can only 
continue recording diagnostic data when in the “synchronous 
magnet mode or during the first 10 cycles of the “auto” 
magnet mode. 
 
2.19  Interrogating and/or Starting 
Statistics 
The recorded diagnostic data (saved data contents of the 
pacemaker) are always read (transmitted through interrogation) 
at the beginning of the follow-up and saved to the programmer.  
This allows relevant data to be displayed on the programmer at 
anytime. 
All statistics can be reset with the “Clear Statistics” function and 
the “Transmit” function in the “Parameter” selection menu.  The 
start time and duration are automatically saved for all statistics.