86  Stratos LV/LV-T Technical Manual   
2.6 Antitachycardia 
Functions: 
The antitachycardia functions include: 
•  Upper basic rate 
• Tachycardia 
mode 
• Tachycardia 
behavior 
• Mode 
Switching 
• PMT 
Management 
•  Preventive Overdrive Pacing 
• Post-AES 
Pacing 
 
2.6.1  Upper Rate and UTR Response 
In atrial-controlled dual chamber modes, the upper tracking 
interval (UTI), along with the atrial refractory period or PMT 
protection window limits the ventricular pacing rate such that it 
will never exceed the programmed upper rate regardless of the 
patient's atrial rate.  
In all triggered modes, the upper tracking interval limits the 
pacing rate that is triggered by sensing.  
N
OTE
Select the upper rate based upon the patient’s tolerance for 
the rate.  The upper rate limit determines the minimal interval 
between a sense or pace event and the subsequent atrial or 
ventricular pacing event.  A shortening of the pacing interval 
to the upper rate interval may also be initiated at rest (e.g., 
by detection of muscle potentials).  Therefore, for patients 
with increased vulnerability a lower programmed upper rate 
is recommended. 
 
2.7 Wenckebach 
2:1 
Wenckebach behavior or 2:1 behavior is available depending on 
the programming of the atrial refractory period, the PMT 
protection window and the upper tracking interval in the modes 
DDD, DDT/A, VDD, DDT and VDT.