Electrical Zero Check 
Turn the power supply “ON” and allow the flowmeter 30 minutes to warm up.  Stop all flow through the 
transducer and check the electrical zero. 
CAUTION: 
DO NOT ASSUME THAT ALL METERING VALVES WILL COMPLETELY 
SHUT OFF FLOW.  EVEN A SLIGHT LEAKAGE THROUGH A VALVE WILL CAUSE AN 
INDICATION ON THE METER WHICH WILL FALSELY APPEAR TO BE A ZERO SHIFT. 
If necessary, adjust the “ZERO” potentiometer located on the front panel of the power supply, until the 
meter indicates zero.