Performance Tests
V. Polarization Dependent Loss (PDL): Optional
C A U T I O N
The patchcord from the source to the polarization controller - with the 
isolator - must not move during and between all measurements.
The patchcords between the polarization controller and the optical 
head must not move from the beginning of the reference measurements 
until these are finished. 
3. Zero the 8153A.
a. Ensure that the laser source is switched off.
b. Press M
ENU
 to change the Measure Mode.
c. Press Z
ERO
 and wait while zeroing. 
4. Set up the laser source.
a. Set the laser source to 1550 nm (nominal), switch the laser 
on, and allow 5 minutes for the laser to settle.
b. Note the actual wavelength in the test record. 
5. Set up the power meter. 
a. Set the power meter to the actual wavelength.
Press P
ARAM
 until the wavelength is displayed, then use the 
modify cursor keys to set the actual wavelength.
b. Set the averaging time to 100 ms.
Press P
ARAM
 until the averaging time is displayed, then use 
the modify cursor keys to set the averaging time to 100 ms.
c. Set the display to W.
Press 
D
B
M
/W
6. Set the polarization filter of the 8169A to maximize the signal.
a. Reset the position of all plates.
Press H
OME
 on the polarization controller.
b. Select the polarization filter.
You may need to press P
OS
 and/or Pol if the filter is not