Agilent Technologies 53152A Manuale Utente

Pagina di 142
Chapter 1  Getting Started
Operating the Counter
Operating Guide
1-17
1
Measuring Frequency
The following diagram shows the basic sequence to use to make a 
frequency measurement using Channel 1. This example assumes that the 
Counter is on and has completed the Self Test. For the purposes of this 
example, use the 10 MHz reference output on the Counter’s rear panel as 
a signal source for input to Channel 1.
The same procedure applies to making a basic frequency measurement on 
Channel 2. However, since Channel 2 is automatically selected when you 
turn on the Counter, the channel-selection step is unnecessary (unless you 
previously selected Channel 1).
CAUTION
The Channel 2 input path circuits contain sensitive GaAs semiconductors. 
To prevent damage to these components, always adhere to standard ESD 
(Electro-Static Discharge) prevention procedures, and ensure that the 
maximum power specification for this channel (+27 dBm) is not exceeded.
Ch   2
      Freq
Ch   1  
      Freq
Chan
Select
Ch   1
      Freq
CHANNEL 1
1M
 
Ω
DAMAGE
+30 dBm
10 Hz to 125 MHz
Ch   2
      Freq