Service Procedures  
181
Appendix D
Modulation Adjustment
The function generator stores three calibration constants related to amplitude 
modulation depth. The constants are calculated from the adjustment value 
entered. If the calibration procedure is aborted before all setup steps have 
been completed, no calibration constants are stored. 
1. Use a DMM to measure the function generator ACrms output voltage 
for SETUP 30 in the following table. These adjustments use a HIGH 
Z output termination.
2. Send the CAL:VALue <measured Vac> to the function generator. 
3. Send the CAL? command to the function generator to initiate the 
calibration.
4. Repeat steps 1, 2 and 3 for SETUP 31 through 32.
5. Perform the “AM Modulation Depth Verification” on page 172.
Note
NEW CALIBRATION:
 A new calibration (SETUP 85  Rev 4.0) has been 
added to eliminate a small residual error in the AM amplitude system which 
could potentially cause a failure of the AM amplitude verification.  
The new calibration operates just like the other AM calibrations (SETUP 
30, 31, 32) in that the external measurement is AC Vrms with no load. The 
new calibration is not allowed until the other AM gain calibrations (SETUP 
30, 31, 32) are performed. 
The new algorithm is designed such that the calibration should not be 
required again once the function generator has been calibrated at the 
factory. However, if you change any critical analog components which 
determine amplitude in AM modulation, you should perform the calibration 
again.
Table D-14. Modulation Adjustment Setups
Nominal Output
SETUP
FREQUENCY
AMPLITUDE
Adjustment for:
30
1 kHz
3.5 Vrms
0% modulation depth.
31
1 kHz
0.707 Vrms
50% modulation depth.
32
1 kHz
6.36 Vrms
100% modulation depth.