Agilent Technologies E1441A Manuale Utente

Pagina di 206
50
Agilent E1441A Application Information  
Chapter 2
To Phase Lock Multiple
Function Generators
(requires Option 001)
The front-panel 10 MHz Ref Out and 10 MHz Ref In terminals allow you to 
synchronize multiple function generators. The following statements show 
you how to synchronize two function generators at 10 kHz (send the 
commands to both function generators):
APPL:SIN 10E+3, 5.0
Select sine function at 10 kHz
PHAS:ADJ -90 DEG
Set phase offset to -90 degrees
PHAS:REF
Set phase reference to zero
Figure 2-12.