Configuration, Startup, and Operation 
 
 
of Recorder with Display 
12/03 
DR4300 Circular Chart Recorder Product Manual 
89 
4.3.8  Pen Parameters Set Up Group 
Introduction 
The functions in this group are used to configure the pen(s). The procedure for configuring each pen is the 
same. For a 2-pen recorder, the desired input channel is displayed on the left side of the operator interface. 
Press FUNC  key to select channel.  
 
Pen group prompts 
Table 4-6 lists all the function prompts in the pen set up group. 
Press the  SET UP  key until “PEN” appears in the lower display. 
Press FUNC  key to display parameters. 
 
Table 4-6  Pen Parameter Definitions 
Lower Display 
Prompt 
Upper Display  
Range of Setting  
or Selection 
Parameter  
Definition 
PEN IN 
 
 
 
INP   [factory setting] 
REM1 
 
 
REM2 
 
 
OUT* 
SP* 
 
PEN INPUT
—What do you want the pen to record? 
INPUT
—Records the input for given channel. 
REMOTE SWITCH 1 EVENT
—Records the digital input 
event for given channel. (Event toggles pen between 
90 % and 95 % on chart.)  
REMOTE SWITCH 2 EVENT
—Records the digital input 
event for given channel. (Event toggles pen between 
80 % and 85 % on chart.)  
OUTPUT—Records the output for the channel. 
SETPOINT—Records setpoint for given channel. 
*Only selectable if hardware supports outputs and 
function prompt “CONTRL” in Control group is set to 
“ENAB” (enabled).  
CHT HI 
 
–999 to 9999 
 
 [factory setting = 780] 
CHART HIGH RANGE VALUE
—Enter a value that 
corresponds with the chart high range value for  
the pen. (This is the value that prints at the outer edge of 
the chart. If you had a range of 0 °C to –400 °C, –400 
would be the chart high range value.) 
CHT LO 
 
–999 to 9999 
 
 [factory setting = 730] 
CHART RANGE LOW VALUE
—Enter a value that 
corresponds with the chart low range value for the  
pen.