Honeywell DR4300 Manuale Utente

Pagina di 262
DR4300 Circular Chart Recorder 
66 
DR4300 Circular Chart Recorder Product Manual 
12/03 
3.3.2  Preparing the Recorder for Operation 
Introduction 
Before applying power to the recorder, complete these preparation tasks. 
ATTENTION 
Never move the pen arms when the unit is operating unless adjusting the pen alignment as 
described in Section 7.  
  
WARNING 
 
SHOCK HAZARD 
When the unit is powered a potentially lethal shock hazard exists at the AC line connections at TB1 on 
each printed circuit assembly (behing the chart plate). Make sure the unit power is disconnected before 
starting the procedure. More than one switch may be required to de-energize the recorder. 
Failure to observe this precaution can result in exposure to a potentially lethal shock hazard. 
Procedure 
Refer to Figure 3-3 to identify the basic chart plate components and follow the procedure in Table 3-3 t
prepare the recorder for operation. 
Table 3-3  Preparing the Recorder for Operation 
Step Action 
Open the door. 
Pull up on the pen lifter to raise the pen(s) from the chart plate and remove the protective cap 
from each pen tip. Lift the locking clip on the chart hub. 
Slip the new chart under the pen lifter, pen and time index, and press it into place over the 
chart hub. 
Turn the chart until the desired time line on the chart is aligned with the time index on the chart 
plate and pen 1 before closing the locking clip on the hub. Push down on the lifter to return the 
pen(s) to the chart. 
CAUTION: Do not try to rotate the hub. 
   
24213
Time index
Large chart
hub
Pen lifter
 