One-Button Measurement Functions
Meas Control
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
multiple data acquisitions, with multiple trigger events, for one full trigger cycle. The
instrument must have external triggering selected, or the command will be ignored. Use
the TRIGer[:SEQuence]:SOURce EXT command to select the external trigger.
History:
Added in revision A.02.00
Remote Command:
:INITiate[:IMMediate]
Remote Command Notes: See also the *TRG command and the TRIGger subsystem.
Use the [:SENSe]:<meas>:TRIGger:SOURce command to select the
desired trigger. The instrument must be in the single measurement mode.
If :INITiate:CONTinuous is ON then the command is ignored.
Use :FETCh? to transfer a measurement result from memory to the output
buffer. Refer to individual commands in the MEASure subsystem for more
information.
Example:
INIT:IMM
5.2.5
Abort the Sweep or Measurement (Remote Command Only)
Stops any sweep or measurement in progress and resets the sweep or trigger system. A
measurement refers to any of the measurements found in the
MEASURE
menu. If the trigger
conditions are met, another sweep is initiated immediately.
If :INITiate:CONTinuous is off (single measure), then :INITiate:IMMediate will
start a new single measurement.
If :INITiate:CONTinuous is on (continuous measure), a new continuous measurement
begins immediately.
The INITiate and/or TRIGger subsystems contain additional related commands.
History:
Added in revision A.02.00
Remote Command:
:ABORt
Remote Command Notes: In the continuous measurement mode, the
Restart
key is
equivalent to ABORt.
Example:
ABOR