One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
indicate this range is currently inactive.
Key Path:
Meas Setup
Saved State:
All values for all ranges are saved in instrument state.
Factory Preset: For each range, the following settings are selected:
Video BW
(Auto),
Sweep Time
(Auto),
Absolute Limit
=
−50 dBm,
Peak Threshold
=
−90 dBm,
Peak
Excursion
= 6 dB, and
Abs Stop Limit Mode
(Auto) is set to
−50 dBm. The table
below defines the other parameter’s settings for the measurement. The
values for ranges 1 through 5 are based on the W-CDMA (3GPP) BTS
Spurious Emission measurement.
Remote Command:
There is no equivalent remote command.
Table 5-6
Range Table Default Settings
Range
Start Freq (GHz)
Stop Freq (GHz)
Res BW (kHz)
1
1.920
1.980
1200
2
1.894
1.920
510
3
2.100
2.102
100
4
2.175
2.180
100
5
0.800
1.000
4000
6-20
1.5
2.5
Auto