Pressure Systems 9816 Manuale Utente
Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©
Page 92
www.PressureSystems.com
SET/DO OPERATING OPTIONS/FUNCTIONS (Command 'w')
Purpose:
Change a module’s default operating option settings, or invoke special internal
operations or functions.
operations or functions.
Command
“wii[dd[ eeee]]”
‘w’ is the command letter.
‘ii’ is the index field.
‘dd’ is an [optional] hex datum.
‘ eeee’ is an [optional] extra datum
with leading space character.
‘w’ is the command letter.
‘ii’ is the index field.
‘dd’ is an [optional] hex datum.
‘ eeee’ is an [optional] extra datum
with leading space character.
Response
“A”
‘A’ is the acknowledge letter.
‘A’ is the acknowledge letter.
Description: The index field (ii) contains two hex digits that identify the specific option to be
set or function to be performed. The datum field (dd), when present, contains 2
hex digits. A few indexes also require an extra datum field (eeee). Valid
options/functions are listed in the table below (-- marks a missing datum field in
its column, and fourth column shows any 'q' command index that reads same
option):
hex digits. A few indexes also require an extra datum field (eeee). Valid
options/functions are listed in the table below (-- marks a missing datum field in
its column, and fourth column shows any 'q' command index that reads same
option):
ii
dd
Description
'q' read
index
00 --
Execute Internal Self Test.
01 --
Update Internal Thermal Coefficients.
02-06
--
Reserved for factory use
07 --
Store Most Operating Options in nonvolatile flash
memory. See *** Note at end of table.
memory. See *** Note at end of table.
08 --
Store Current Offset Terms in transducers’ non-volatile
memories. See *** Note at end of table.
memories. See *** Note at end of table.
09 --
Store Current Gain Terms in transducers’ non-volatile
memories. See *** Note at end of table.
memories. See *** Note at end of table.
0A 01-10
(1-16)
Set Number of Channels in Module (default =16).
0B
00
01
Enable Automatic Shifting of Calibration Valve during
Calculate and Set Offsets ('h') command (default).
Disable Automatic Shifting of Calibration Valve in 'h'.
User will manually control calibration value.
Calculate and Set Offsets ('h') command (default).
Disable Automatic Shifting of Calibration Valve in 'h'.
User will manually control calibration value.