6.1.2  
Checking Module LED Power-Up Sequence  
 
(1) 
Power to each module should be verified.  Since module power is supplied by the 98RK-
1 Scanner Interface Rack
, make sure that the rack is plugged into the proper type of 
power receptacle. 
 
 
(2) 
 
Turn the rack power supply ON and verify the front panel LED status (for each installed 
scanner) following initial power-up: 
 
● 
PWR LED should remain ON
 
If this LED is not on, all other LEDs will likely be off.  Make sure the module is  
 
properly seated (mated) with the rack and the rack is plugged into a proper power 
 
receptacle.  Do not unplug or reseat a module while rack power is ON. 
 
● 
TX LED should remain OFF
 
Note that any activity of the TX LED during the power-up sequence is an  
 
indication that the Dynamic IP assignment protocol is enabled.  This will typically  
 
occur following the initial BSY LED cycle and continue until an appropriate server  
 
reply is received. 
 
● 
LNK LED should remain ON
 
This LED indicates proper connection to an Ethernet hub or switch.  If this LED is  
 
off, verify that the rack is properly connected to the communications hub or  
 switch. 
 
 
● 
COL LED should remain OFF
 
● 
PRG LED should remain OFF
 
● 
CAL LED should remain OFF
 
● 
BSY  LED should blink 8-16 times at a rate somewhat slower than 1 Hz, delay ON 
 
for approximately 30 seconds and then start to continuously blink at  
 
approximately 100 Hz (rate is dependent on the number of internal averages).  
 
This faster blink rate may simply appear as a dimly-lit LED.  Note that the BSY  
 
LED will not begin its 100 Hz blink rate if the Dynamic IP assignment protocol is 
 
enabled and the module has not received a proper IP address response. 
 
 
Any significant variation from this power-up LED sequence is an indication of a possible rack 
problem or a 9816 microprocessor board (PC-299) error.  If the proper power-up LED sequence 
is not achieved after following the above suggestions, contact the Repair Department at 
Pressure Systems for additional assistance (1-800-678-SCAN).