Constant Voltage Zero Calibration
Note
:
Perform this procedure only after completing Voltage Monitor and Remote Readback Full Scale
Calibration.
a.  Send string "VSET 0; ISET 5; OUT ON".
b.  Connect an external supply to the power supply as shown in Figure 2-2.
c.  Attach the DVM from - S to + S on the rear panel.
d. Adjust A8R40 (CV PROG ZERO) to 0 
±120µV.
Current Monitor Zero Calibration
a.  Send string "VSET 0; ISET 0; OUT OFF''.
b.  Connect a short across power supply output terminals.
c.  Attach the DVM from   M to IM on the rear panel.
d. Allow several minutes (3 or more) to ensure thermal settling.
e. Adjust A2R8 (I-MON ZERO) to 
0 ±100 µV.
Constant Current Zero Calibration
a.  Connect the test setup shown in Figure 2-3.
b.  Send string ''VSET 5; ISET 0; OUT ON''.
c.  Allow several minutes (3 or more) to ensure thermal settling.
d.  Adjust A8R29 (CC PROG ZERO) to:
0V 
±0.6mV (6030A, 6035A)
0V 
±8.0mV (6031A)
0V 
±1.7mV (6032A)
Figure 2-3. CC Zero Calibration Setup